HomeMy WebLinkAboutDRC-2025-002811Sample Container, Volumes, and Preservation Requirements
Level:
Work
InstructionDocument number:STL-S-SR-WI59743
Old Reference:
ST-QA-0045
Version:
5.1 Organisation level:
4-Business Unit
Approved by: ORD0
Effective Date: 30-
MAY-2025
Document users:
STL_CSO Project Managers All, STL_Sample Mgmt All
Responsible:
STL_QA All
The information contained herein is of a highly confidential and proprietary nature. Eurofins Environment Testing specifically prohibits the
dissemination, copy, disclosure, transfer, or modification of this information without the express written approval of Eurofins Environment Testing
St. Louis QA.
Metals
Analysis Method Matrix Volume1 / Container Preservative Hold Time2 QC (Default)
Mercury 7470 Water 3 0mL / P, G H NO3 (pH <2) 28 days MS/MSD
7471 Solid 0.6g / G Cool, 0-6° C 28 days M S/MSD
Metals
200.7
200.8
6010
6020
Water 50mL / P, G H NO3 (pH <2) 180 days M S/MSD
Solid 0.5g / G Cool upon receipt, 0-6° C 180 days MS/MSD
Radiochemistry
Analysis Method Matrix Volume1
/ Container Preservative Hold Time2 QC(Default)
Carbon-14
(C-14) E ERF C-01
Water
-----
Solid
500mL / P, G
------------
5g / P, G
DO NOT PRESERVE
------------
DO NOT DRY
N/A MS/Dup
Chlorine-36
(Cl-36)
In-house Water
------
50mL / P, G
------------
DO NOT PRESERVE
------------
N/A Duplicate
Always check on-line for validity.
Solid 2g / P NONE
Gross Alpha/Beta3 900.0
9310
Water
------
Solid
200mL / P, G
-----------
1g / P, G
HNO3 (pH <2)
------------
None
N/A MS/Dup
Gross Alpha
Co-precip SM 7110C Water 500mL / P, G HNO3 (pH <2) N/A MS/Dup
Iodine-129 (I-129)
(Carrier Analysis)
Liquid Scint
------------
901.1
GA-01-R
MOD
Water
------
Solid
2L / P, G
----------
650g / P, G
DO NOT PRESERVE
------------
DO NOT DRY
28 days Duplicate
Iodine-131
(I-131)
901.1
GA-01-R
MOD
Water
------
Solid
1L / P, G
------------
650g / P, G
DO NOT PRESERVE
------------
DO NOT DRY
28 days Duplicate
Iron-55
(Fe-55)
(Carrier Analysis)
In-house
Water
------
Solid
500mL
------
5g
HNO3 (pH <2)
------------
None
N/A Duplicate
Gamma Scan
901.1
HASL 300
GA-01-R
Water
------
Solid
1L / P, G
------------
650g / P, G
HNO 3 (pH <2)
------------
None
N/A Duplicate
Americium-241/Curium-243
244 (Tracer Analysis)
HASL 300
A-01-R
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Americium-243/Curium-
245,246,247,248
(Tracer Analysis)
HASL 300
A-01-R
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Isotopic Plutonium
(Tracer Analysis)
HASL 300
A-01-R
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Isotopic Thorium
(Tracer Analysis)
HASL 300
A-01-R
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Isotopic Uranium
(Tracer Analysis)
HASL 300
A-01-R
HASL 300
U-02 -RC
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
-------------
None
N/A Duplicate
Lead-210
(Pb-210)
(Carrier Analysis )
EERF PB-01
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Nickel-59/63
(Ni-59/63)
(Carrier Analysis )
In-house
Water
------
Solid
500mL / P, G
------------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Polonium-210
(Po-210)
(Tracer Analysis)
HASL 300
Po-01-RC
Water
------
Solid
500mL / P, G
------------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Radon-222 SM 7500-Rn
EPA 913
Water
------
Bottled
Water
40mL / G vial Insulate
4 days
--------------
4 days once
opened
Duplicate
Radium-226
(Ra-226)
(Carrier Analysis)
903.0
9315
Water
------
Solid
1L / P, G
-------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Radium-228
(Ra-228)
(Carrier Analysis)
904.0
9320
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Total Alpha Radium
(Ra-223/224/226)
(Carrier Analysis)
903.0
9315
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Selenium-79 In-house
Water
------
Solid
500mL / P, G
------------
1g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Strontium-89
(Sr-89)
(Carrier Analysis)
905.0
HASL 300
Sr-01-RC
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Strontium-90
(Sr-90)
(Carrier Analysis)
905.0
HASL 300
Sr-02-RC
HASL 300
Sr-03-RC
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Strontium-90 by Extraction
(Sr-90)
ASTM D5811 Water 1L / P, G HNO3 (pH <2)N/A Duplicate
(Carrier Analysis)
------
ASTM C1507
------
Solid
------
5g / P, G
------------
None
Total Beta Strontium
(Carrier Analysis)
905.0
HASL 300
Sr-03-RC
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Total Beta Strontium by
Extraction
(Carrier Analysis)
ASTM D5811
-------
ASTM C1507
Water
------
Solid
1L / P, G
------
5g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Technetium-99
(Tc-99)
(Tracer Analysis)
HASL 300
Tc-02-RC
Water
------
Solid
1L / P, G
------
10g / P, G
HNO3 (pH <2)
------------
None
N/A Duplicate
Tritium (H-3)
906.0
HASL 300
H-04-RC
Water
------
Solid
120mL / G4
------------
100g / G4
DO NOT PRESERVE
------------
DO NOT DRY
N/A Duplicate
1Sample volumes are based on dry weights; volumes need to be increased if soil is wet/moist. For samples requiring Matrix QC, 3 times the
volume is required. For normal samples, 2 or more times the volume may be required for re-extracts/digestions.
2 For a test method with a maximum holding time measured in days, the holding time shall be tracked by the day. The sampling date is day 0.
3Gross Alpha MDA is achievable only when solids are less than 500 ppm.
4 Glass is the preferred container type due to the slight potential for tritium to leach through a plastic container over time. However this is
unlikely to be a significant factor, and a glass container is not required. Plastic containers are acceptable.
Container Key
P - Plastic
G - Glass
This table should be used for guidance only. If there is any doubt or question, please consult the appropriate group leader. Some
analyses may be able to be combined, such as the anions, with no additional volume required. Check with group leader if there
are changes in the method or the manner in which the lab performs an analysis.
End of document
Version history
Version Approval Revision information
4 10.SEP.2024
5 24.JAN.2025 Removed "do not dry" note from Tc-99 Solids
5.1 30.MAY.2025 Fixed formatting.