RE: First Quarter 2024 Calibration Error Results
Clean Harbors Aragonite, LLC
EPA ID# UTD 981552 177
Dear Mr. Hansen,
In accordance with Attachment 15 Section 7.2 of the Clean Harbors Aragonite, LLC RCRA Permit,
attached are results of the First Quarter 2024 Calibration Error Testing, which indicate that the continuous
emission monitors (CEM) are in proper working condition.
I hereby certify under penalty of law that this document was prepared under my direct supervision in
accordance with a system designed to assure that qualified personnel properly gather and evaluate the
information submitted. Based upon my inquiry of the person or persons who manage the system, or those
persons directly responsible for gathering the information submitted is, to the best of my knowledge and
belief, true, accurate and complete. I am aware that there are significant penalties for submitting false
information, including the possibility of fine and imprisonment for known violations.
Should you have any questions regarding this matter, please contact me at the number listed below.
William Simmons
Facility GM III
Clean Harbors Aragonite, LLC
l 1600 North Aptus Road
PO Box 1339
Grantsville, Utah 84029-1339
(o) 435.844.8351
(c) 870.310.6029
simmons. wi I liam@cleanharbors.com
www .cleanharbors.com
• ~
tleJip~r_bprJ
-••wa.ACJL.QV.a.xu ... zua._
MEMORANDUM FOR RECORD
DATE: March 28, 2024
SUBJECT: CALIBRATION ERROR TESTING
CLEAN HARBORS ARAGONITE, LLC
The calibration error testing for the #1 CEM was completed for the carbon monoxide (low and
high) analyzers, the oxygen monitor, total hydrocarbon analyzer, and the carbon dioxide monitor.
The results are given in the table below:
Instrument Date of Test Error Error mid Error high Pass criteria zero/low
CO Io 2/29/2024 1.09 0.04 0.60 5% of span
CO hi 2/21/2024 0.43 0.20 0.12 5% of span
02 2/20/2024 0.07 0.04 0.03 0.5% volume
THC 3/05/2024 0.10 0.35 0.30 5% of span
CO2 2/21/2024 0.18 0.11 0.06 0.5% volume
The calibration error testing for the #2 CEM was completed for the carbon monoxide (low and
high) analyzers, the oxygen monitor, total hydrocarbon analyzer, and the carbon dioxide monitor.
The results are given in the table below:
Instrument Date of Test Error zero/low Error mid Error high Pass criteria
CO lo 2/29/2024 0.16 0.06 0.40 5% of span
CO hi 2/27/2024 0.51 0.57 0.11 5% of span
02 2/20/2024 0.05 0.05 0.14 0.5% volume
THC 3/05/2024 0.12 0.30 0.86 5% of span
CO2 2/21/2024 0.16 0.12 0.01 0.5% volume
The procedure followed was the calibration error procedure as described in 40 CFR part 266,
Appendix IX. Attached to this memo are copies of the field data sheets.
Date of Test: 29 Feb 2024
Quarter: 1
Monitor Type: Low CO
Manufacturer: Servomex
Serial Number: 653204
Run Number Calibration
Value
1 -zero 19.78
2-mid 70.48
3 -hi h 150.6
4-mid 19.78
5 -zero 70.48
6-hi h 150.6
7 -zero 19.78
8-mid 70.48
9-hi h 150.6
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Dave Devey ----'-'--------
mA Value
5.76
9.64
16.15
9.65
5.76
16.13
5.75
9.63
16.15
Monitor
Response
22.0
70.50
151.87
70.63
22.0
151.63
21.88
70.38
151.87
Mean
Difference =
Calibration
Error% of
Span=
CEM # --'--I _________ _
Span: 0 -200 PPM
Zero/Low 1 Mid 1 High 1
2 2 ' 2 -
2.18 .09 1.19
3 3 3
1.09 .04 .60
Test Times (start and finish): 11:43 --12:00
Cylinder ID No . CC359903
Certification Date: 13Jul 2020
Expiration Date: 13 Jul 2028
Cylinder ID No. CC480934
Certification Date: 23 Dec 2019
Expiration Date: 23 Dec 2027
Cylinder ID No. ALM024621
Certification Date : 14 Aug 2020
Expiration Date: 14Aug 2028
Comments:
1 Absolute Difference= I Calibration value -
Monitor ,Response I
2 Mean Difference = Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 29 Feb 2024
Quarter: 1
Monitor Type: low CO
Manufacturer: Servomex
Serial Number: 653305
Run Number Calibration
Value
1 -zero 19.78
2 -mid 70.48
3 -hi h 150.6
4-mid 19.78
5 -zero 70.48
6 -hi h 150.6
7 -zero 19.78
8-mid 70.48
9-h i h 150.6
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Dave Devey ____ _,__ _____ _
mA Value
5.55
9.64
15.98
9.62
5.56
16.0
5.56
9.63
15.97
Monitor
Response
19.38
70.50
149.75
70.25
19.50
150.0
19.50
70.38
149.63
Mean
Difference =
Calibration
Error% of
Span=
CEM # -'2=-------------
Span: 0 -200 PPM
Zero/Low 1 Mid 1 High 1
.97
2 2
.32 .12 .81
3 3 3
.16 .06 .40
Test Times (start and finish): 15:20 --15:31
Cylinder ID No. CC359903
Certification Date: 13Jul 2020
Expiration Date : 13 Jul 2028
Cylinder ID No. CC480934
Certification Date: 23 Dec 2019
Expiration Date: 23 Dec 2027
Cylinder ID No. ALM024621
Certification Date: 14 Aug 2020
Expiration Date: 14 Aug 2028
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2Mean Difference = Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Date of Test: 21 Feb 2024 Test Technicians: Nathan Haney ------'-------
Quarter: 1
Monitor Type: _H_,ig"-h_C_O ________ _
Manufacturer:
Serial Number:
Run Number
2-mid
3 -hi h
4-mid
5 -zero
6-hi h
7 -zero
8-mid
9-hi h
Servomex
653204 ------------
Calibration
Value
1251.
3520.
7495.
1251.
3520.
7495.
1251.
3520.
7495.
mA Value
5.93
9.66
15.98
9.68
5.93
15.98
5.94
9.65
15.96
Monitor
Response
1206.25
3537.50
7487.50
3550.0
1206.25
7487.50
1212.50
3531.25
7475.0
Mean
Difference =
Calibration
Error% of
Span=
Dave Devey
CEM # -=--1 _________ _
Span: 0 -10000 PPM
Zero/Low1 Mid1 High 1
2 2 • .. 2
42.67 19.58 11.67
3 3 3
.43 .20 .12
Test Times (start and finish): 15:55 --16:08
Cylinder ID No. CC266951 ------------
Ce rt i fi cation Date: 4 Oct 2022
Expiration Date : 4 Oct 2030 ------------
Cylinder ID No. CC213578 ------------
Certification Date: 1 Sep 2023
Expiration Date : 1 Sep 2031 __ ..,__ ________ _
Cylinder ID No. CC330683 ------------
Ce rt i fi cation Date: 1 Sep 2023
Expiration Date: 1 Sep 2031 __ ..,__ ________ _
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2Mean Difference = Sum of Absolute Differences/ 3
3 Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 27 Feb 2024
Quarter: 1
Monitor Type: High CO
Manufacturer: Servomex
Serial Number: 653305
Run Number Calibration
Value
1 -zero 1251.
2-mid 3520.
3 -hi h 7495.
4-mid 1251.
5 -zero 3520.
6-hi h 7495.
7 -zero 1251.
8-mid 3520.
9-hi h 7495.
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Nathan Haney ------'------
mA Value
6.08
9.72
16.01
9.73
6.08
16.0
6.09
9.72
16.02
Monitor
Response
1300.0
3575.0
7506.25
3581.25
1300.0
7500.0
1306.25
3575.0
7512.50
Mean
Difference =
Calibration
Error% of
Span=
Dave Devey
CEM# _2 _________ _
Span: 0 -10000 PPM
Zero/Low 1 Mid 1 High 1
51.08 57.08 11.25
3 3 3
.51 .57 .11
Test Times (start and finish): 14:52 --15:03
Cylinder ID No . CC266951
Certification Date: 4 Oct 2022
Expiration Date : 4 Oct 2030
Cylinder ID No. CC213578
Certification Date: 1 Sep 2023
Expiration Date: 1 Sep 2031
Cylinder ID No . CC330683
Certification Date: 1 Sep 2023
Expiration Date: 1 Sep 2031
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2 Mean Difference= Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Date of Test: 20 Feb 2024 ------------Test Technicians: Nathan Haney ____ _____;..!,...._ ____ _
Quarter: 1 -------
Monitor Type: 02 ------------
Manufacturer: Servomex ------------Serial Number: 4021 ------------
Run Number Calibration mA Value
Value
1 -zero .996 4.68
2-mid 9.033 9.80
3 -hi h 15.04 13.63
4-mid .996 9.83
5 -zero 9.033 4.68
6-hi h 15.04 13.65
7 -zero .996 4.68
8-mid 9.033 9.77
9-hi h 15.04 13.65
Monitor
Response
1.06
9.06
15.05
9.11
1.06
15.08
1.06
9.02
15.08
Mean
Difference =
Calibration
Error% of
Span=
CEM # _l:__ _________ _
Span: 0-25 %
Zero/Low 1 Mid 1 High 1
2 2 .2
.07 .04 .03
3 3 3
.27 .16 .11
Test Times (start and finish}: 13:15 --13:33
Cylinder ID No. EB0122991
Certification Date: 1 Nov 2021
Expiration Date: 1 Nov 2029
Cylinder ID No. CC47205
Certification Date: 14Jul2020
Expiration Date: 14Jul2028
Cylinder ID No. CC151392
Certification Date : 31 Aug 2020
Expiration Date: 31 Aug 2028
Comments:
1 Absolute Difference= I Calibration value -
Monitor Response I
2 Mean Difference= Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span} x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range}
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Date of Test: 20 Feb 2024 ------------Test Technicians: Nathan Haney ------=--------
Quarter: _1 _____ _
Monitor Type: _0_2 __________ _
Manufacturer: Servomex
Serial Number: 652525 ------------
Run Number Calibration mA Value
Value
1 -zero .996 4.60
2-mid 9.033 9.81
3 -hi h 15.04 13.71
4-mid .996 9.81
5 -zero 9.033 4.61
6 -hi h 15.04 13.72
7 -zero .996 4.61
8-mid 9.033 9.81
9-hi h 15.04 13.72
Monitor
Response
.94
9.08
15.17
9.08
.95
15.19
.95
9.08
15.19
Mean
Difference =
Calibration
Error% of
Span=
CEM # ....:2=--------------
Span: 0-25 %
Zero/Low 1 Mid 1 High 1
.15
2 2 2 .
.05 .05 .14
3 3 3
.19 .18 .57
Test Times (start and finish): 14:18 --14:34
Cylinder ID No. EB0122991
Certification Date: 1 Nov 2021
Expiration Date: 1 Nov 2029
Cylinder ID No. CC47205
Certification Date : 14Jul2020
Expiration Date: 14Jul2028
Cylinder ID No . CC151392
Certification Date : 31 Aug 2020
Expiration Date: 31 Aug 2028
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2Mean Difference= Sum .of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 5 Mar 2024
Quarter: 1
Monitor Type: THC
Manufacturer: Thermo Scientific
Serial Number: 1203299855
Run Number Calibration
Value
1 -zero .0
2-mid 35.04
3 -hi h 75.38
4-mid .0
5 -zero 35 .04
6 -hi h 75 .38
7 -zero .0
8-mid 35.04
9 -hi h 75.38
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Nathan Haney _____ .,__ ____ _
mA Value
4.0
9.58
16.01
9.69
4.01
16.04
4.04
9.55
15.99
Monitor
Response
.00
34.88
75.06
35.56
.06
75.25
.25
34.69
74.94
Mean
Difference =
Calibration
Error% of
Span=
CEM # ---=-1 _________ _
Span: 0 -100 PPM
Zero/Low 1 Mid 1 High 1
2 2 ' 2
.10 .35 .30
3 3 3
.10 .35 .30
Test Times (start and finish): 14:21 --14:46
Cylinder ID No. ALM052593
Certification Date: 23 Aug 2023
Expiration Date: 23 Aug 2031
Cylinder ID No. EB0122971
Certification Date: 3 Oct 2022
Expiration Date: 3 Oct 2030
Cylinder ID No. CC189252
Certification Date : 6 Sep 202 3
Expiration Date: 6 Sep 2031
Comments:
1 Absolute Difference= I Calibration value -
Monitor Response I
2 Mean Difference = Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 5 Mar 2024
Quarter: 1
Monitor Type: THC
Manufacturer: Thermo Scientific
Serial Number: 1203299856
Run Number Calibration
Value
1 -zero .0
2-mid 35.04
3 -hi h 75.38
4-mid .0
5 -zero 35.04
6 -hi h 75.38
7 -zero .0
8-mid 35.04
9 -hi h 75.38
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
I
J
Test Technicians: Nathan Haney _____ .,__ ____ _
mA Value
4.01
9.55
15.93
9.62
4.01
15.92
4.04
9.53
15.92
Monitor
Response
.06
34.69
74.56
35.13
.06
74.50
.25
34.56
74.50
Mean
Difference =
Calibration
Error% of
Span=
CEM # _:2::...,_ _________ _
Span: 0 -100 PPM
Zero/Low 1 Mid 1 High 1
.12 .30 .86
3 3 3
.12 .30 .86
Test Times (start and finish): 15:00 --15:23
Cylinder ID No. ALM052593
Certification Date : 23 Aug 2023
Expiration Date: 23 Aug 2031
Cylinder ID No. EB0122971
Certification Date: 3 Oct 2022
Expiration Date: 3 Oct 2030
Cylinder ID No. CC189252
Certification Date: 6 Sep 2023
Expiration Date: 6 Sep 2031
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2 Mean Difference= Sum of Absolute Differences/ 3
3 Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 21 Feb 2024
Quarter: 1
Monitor Type: Co2
Manufacturer: Servomex
Serial Number: 653204
Run Number Calibration
Value
1 -zero 2.007
2-mid 7.006
3 -hi h 15.14
4-mid 2.007
5 -zero 7.006
6-hi h 15.14
7 -zero 2.007
8-mid 7.006
9-hi h 15.14
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Nathan Haney -----~-----
mA Value
5.75
9.67
16.09
9.74
5.75
16.05
5.75
9.68
16.05
Monitor
Response
2.19
7.09
15.11
7.18
2.19
15.06
2.19
7.10
15.06
Mean
Difference =
Calibration
Error% of
Span=
CEM # --=-----------
Span: 0-20 %
Zero/Low 1 Mid 1 High 1
.08
2 2
.18 .11 .06
3 3 3
.90 .57 .30
Test Times (start and finish): 11:18 --11:33
Cylinder ID No. XC031310B
Certification Date: 30 Aug 2016
Expiration Date: 30Aug 2024
Cylinder ID No . EB0123063
Certification Date: 23 Dec 2019
Expiration Date: 23 Dec 2027
Cylinder ID No. EB0106597
Certification Date: 5 Mar 2019
Expiration Date: 5 Mar 2027
Comments:
1 Absolute Differ~nce = I Calibration value -
Monitor Response I
2Mean Difference = Sum of Absolute Differences/ 3
3 Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%
Date of Test: 21 Feb 2024
Quarter: 1
Monitor Type: Co2
Manufacturer: Servomex
Serial Number: 653305
Run Number Calibration
Value
1 -zero 2.007
2-mid 7.006
3 -hi h 15.14
4-mid 2.007
5 -zero 7.006
6-hi h 15.14
7 -zero 2.007
8-mid 7.006
9-hi h 15.14
Continuous Emission Monitor
Calibration Error Test
Field Data Sheet
Test Technicians: Nathan Haney ------'------
mA Value
5.73
9.69
16.12
9.72
5.74
16.12
5.74
9.69
16.12
Monitor
Response
2.16
7.11
15.15
7.15
2.18
15.15
2.18
7.11
15.15
Mean
Difference =
Calibration
Error% of
Span=
CEM # -'2~----------
Span: 0-20 %
Zero/Low 1 Mid 1 High 1
.01
.2 2 . 2
.16 .12 .01
3 3 3
.82 .59 .OS
Test Times (start and finish): 12:26 --12:35
Cylinder ID No. XC031310B
Certification Date: 30 Aug 2016
Expiration Date: 30Aug 2024
Cylinder ID No. EB0123063
Certification Date : 23 Dec 2019
Expiration Date: 23 Dec 2027
Cylinder ID No. EB0106597
Certification Date : 5 Mar 20 19
Expiration Date: 5 Mar 2027
Comments:
1 Absolute Difference = I Calibration value -
Monitor Response I
2 Mean Difference = Sum of Absolute Differences/ 3
3Calibration Error= (Mean Difference/ Span) x 100%
3
CO pass/fail
</= 5% of span
</=10 ppm (lo range)
</=100 ppm(hi range)
2
02 and Co2 pass/fail
Criteria
Co2 and 02 </= 0.5%