Re: Chevron Marketing Terminal (Chevron) – Pretest Protocol for Volatile Organic
Compounds (VOC) Relative Accuracy Test – Salt Lake County
The Utah Division of Air Quality (DAQ) reviewed Chevron’s pretest protocol submitted by email
on October 21, 2024. The proposed test shall be performed under the following conditions:
Chevron’s monitoring system to be certified:
Point Source Channel Manufacturer Model # Monitor SN
Primary VRU VOC Infrared Technologies IR-8400DC 4303
Secondary VRU VOC Infrared Technologies IR-8400DC 5235
Alliance Technical Group will conduct a relative accuracy/performance specification test on
Chevron’s continuous monitoring system.
Relative accuracy must be determined in the units of the emission standard or equivalent units of
the emission standard, i.e., tons/year shall be tested in pounds/hour or lb Hg/MM ton clinker
30-DRA shall be tested as lb Hg/MM ton clinker.
The DAQ understands that Alliance Source Testing’s staff will conduct testing as follows:
Test Date Point Source Parameter EPA Reference Method
December 26, 2024 Primary VRU NMOC 25B
December 27, 2024 Secondary VRU NMOC 25B
1 , " / Û Ü Û Ù Û Ý
DAQC-1079-24
Page 2
• Reference Method 25B - Determination of Total Gaseous Non-Methane Organic
Concentrations Using a Non-dispersive Infrared Analyzer as outlined in 40 CFR 60
Appendix A. Zero drift and calibration drift checks are required at the beginning
and end of each run.
• Deviations - The Director or the representative of the Director must be notified of
and approve deviations of the Reference Method test. Any deviation from these
conditions without approval from the DAQ may constitute rejection of these tests.
Acceptance of a protocol does not relieve the owner/operator and the testing
contractor from strict adherence to all applicable EPA methods, DAQ policies,
Utah Air Quality Rules (UAQR), and methods approved by the Director. Any
deviation from EPA methods, DAQ policies, UAQR, and methods approved by the
Director must be addressed separately and express written consent given prior to
commencement of testing.
• Field data (Point Source emission data and test data) shall be given to the DAQ in a
timely manner for review.
• The Director will determine the relative accuracy of each monitoring system based
on Point Source continuous emission monitoring data and test data acquired by the
staff member of the DAQ during the test.
• All test reports must be submitted to the Director not later than 60 days after
completion of the test.
• The test report shall include all raw calibration data, raw emission data with date,
time stamps.
Relative accuracy test reports must contain:
1. Point source continuous monitor, channel, manufacturer, and serial number,
2. Raw stack test data and continuous monitor data with date and time stamps.
3. Emission data reported in concentration (ppm or %) and units in the applicable
emission limit.
The DAQ requires that all test reports include a statement signed by a responsible official
certifying that:
1. Testing was conducted while the Point Source was operating at the rate and/or
conditions specified in the applicable approval order, operating permit, or federal
regulation.
DAQC-1079-24
Page 3
2. During testing, the Point Source combusted fuels, used raw materials, and
maintained process conditions representative of normal operations, and operated
under such other relevant conditions specified by the Director.
3. Based on information and belief formed after reasonable inquiry, the statements
and information contained in the report are true, accurate, and complete.
Should you have any questions concerning this matter, contact me at (801) 536-4438 or by email
at rleishman@utah.gov.
Sincerely,
Rob Leishman, Environmental Scientist
Major Source Compliance Section
RL:rh
cc: Salt Lake County Health Department
Alliance Technical Group
* $ . # ( ) - — / A A v A ? A C @ @ w D @ ˜
Rob Leishman Jr