HomeMy WebLinkAboutDSHW-2024-006979Clean Harbors Aragonite, LLC . 11600 N Aptus Road P.O. Box 1339 Aragonite, UT 84029 June 27, 2024 Submitted via email to dwmrcsubmit@utah.gov Mr. Douglas J. Hansen, Director Division of Waste Management & Radiation Control Department of Environmental Quality 195 North 1950 West P.O. Box 144880 Salt Lake City, UT 84114-4880 RE: Second Quarter 2024 Calibration Error Results Clean Harbors Aragonite, LLC EPA ID# UTD 981 552177 Dear Mr. Hansen, In accordance with Attachment 15 Section 7.2 of the Clean Harbors Aragonite, LLC RCRA Permit, attached are results of the Second Quarter 2024 Calibration Error Testing, which indicate that the continuous emission monitors (CEM) are in proper working condition. I hereby certify under penalty of law that this document was prepared under my direct supervision in accordance with a system desi.gned to assure that qualified personnel properly gather and evaluate the information submitted. Based upon my inquiry of the person or persons who manage the system, or those persons directly responsible for gathering the infonnatioo submitted is , to the best of my knowledge and belief, true, accurate and complete . r am aware that there are significant penalties for submitting false information, including the possibility of fine and imprisonment for known violations. Should you have any questions regarding this matter, please contact me at the number listed below. William Simmons Facility GM III Clean Harbors Aragonite, LLC 11600 North Aptus Road PO Box 1339 Grantsville, Utah 84029-1339 (o) 435.844.8351 (c) 870.310.6029 simmons.william@cleanharbors.com www.cleanharbors.com DSHW-2024-006979 (J~~ar~bPl(S· ZU...<WWlWJL.UJLWKU- MEMORANDUM FOR RECORD DATE: June 27, 2024 SUBJECT: CALIBRATION ERROR TESTING CLEAN HARBORS ARAGONITE, LLC The calibration error testing for the # 1 CEM was completed for the carbon monoxide (low and high) analyzers, the oxygen monitor, total hydrocarbon analyzer, and the carbon dioxide monitor. The results are given in the table below: Instrument Date of Test Error Error mid Error high Pass criteria zero/low CO lo 5/9/2024 0.38 0.20 1.12 5% of span CO hi 5/9/2024 0.05 0.42 0.10 5% of span 02 5/16/2024 0.03 0.04 0.11 0.5% volume THC 5/10/2024 0.10 0.33 0.86 5% of span CO2 5/16/2024 0.20 0.14 0.02 0.5% volume The calibration error testing for the #2 CEM was completed for the carbon monoxide (low and high) analyzers, the oxygen monitor, total hydrocarbon analyzer, and the carbon dioxide monitor. The results are given in the table below: Instrument Date of Test Error zero/low Error mid Error high Pass criteria CO lo 5/9/2024 0.26 0.26 0.10 5% of span CO hi 5/9/2024 0.55 0.40 0.08 5% of span 02 5/16/2024 0.23 0.19 0.05 0.5% volume THC 5/15/2024 0.08 0.32 0.72 5% of span CO2 5/16/2024 0.20 0.14 0.01 0.5% volume The procedure followed was the calibration error procedure as described in 40 CFR part 266, Appendix IX. Attached to this memo are copies of the field data sheets. Date of Test: 9 May 2024 Quarter: fl 1.. Monitor Type: Low CO Manufacturer: Servomex Serial Number: 653204 Run Number Calibration Value 1 -zero 19.78 2-mid 70.48 3 -hi h 150.6 4-mid 19.78 5 -zero 70.48 6-hi h 150.6 7 -zero 19.78 8-mid 70.48 9 -hi h 150.6 vUUUllUUU:) .L,.UH:):)1UH 1V.lU.UHV1 Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney _____ ..,__ _____ _ mA Value 5.64 9.61 16.23 9.61 5.64 16.21 5.65 9.60 16.24 Monitor Response 20.50 70.13 152.88 70.13 20.50 152.63 20.63 70.0 153.0 Mean Difference = Calibration Error% of Span= CEM # --=---1 _________ _ Span: 0 -200 PPM Zero/Low 1 Mid 1 High 1 2 2 2 .76 .40 2.23 3 3 3 .38 .20 1.12 Test Times (start and finish): 08:09 -08:19 Cylinder ID No. CC359903 Certification Date : 13 JuJ 2020 Expiration Date : 13 Jul 2028 Cylinder ID No. CC480934 Certification Date: 23 Dec 2019 Expiration Date: 23 Dec 2027 Cylinder ID No. ALM024621 Certification Date : 14 Aug 2020 Expiration Date: 14 Aug 2028 Comments: 1 Absolute Difference= I Calibration value - Monitor Response I 2 Mean Difference= Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fa ii </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: 9 May 2024 Quarter: A'.' :l. Monitor Type: Low CO Manufacturer: Servomex Serial Number: 653305 Run Number Calibration Value 1 -zero 19.78 2-mid 70.48 3 -hi h 150.6 4-mid 19.78 5 -zero 70.48 6-hi h 150.6 7 -zero 19.78 8-mid 70.48 9-h~h 150.6 '"-.,,V.1.1L11J.UVU.:) L.1.lJ.J..:).:JJ.V.1.1 .lV.lU.1.llLV.l Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney ------'-------- mA Value 5.62 9.67 16.07 9.69 5.62 16.07 5.63 9.68 16.05 Monitor Response 20.25 70.88 150.88 71.13 20.25 150.88 20.38 71. 0 150.63 Mean Difference= Calibration Error% of Span= CEM# 2 ------------ Span: 0 -200 PPM Zero/Low 1 Mid 1 High 1 .51 .52 .19 3 3 3 .26 .26 .10 Test Times (start and finish): 08:26 --08:37 Cylinder ID No. CC359903 Certification Date: 13Jul2020 Expiration Date: 13Jul2028 Cylinder ID No. CC480934 Certification Date: 23 Dec 2019 Expiration Date: 23 Dec 2027 Cylinder ID No. ALM024621 Certification Date : 14 Aug 2020 Expiration Date: 14 Aug 2028 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference= Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: 9 May 2024 Quarter: 'L-UHUHUUU.:l .Ll1111.:l.:l1U11 1V1.U111LU1 Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney _____ _,__ _____ _ Monitor Type: High co ---=----------- Manufacturer: Servomex Serial Number: 653204 ------------ Run Number Calibration mA Value Value 1 -zero 1251. 5.98 2-mid 3520. 9.69 3 -hi h 7495. 15.98 4-mid 1251. 9.71 5 -zero 3520. 6.0 6-hi h 7495. 15.98 7 -zero 1251. 6.0 8-mid 3520. 9.70 9-hi h 7495. 15.97 Monitor Response 1237.50 3556.25 7487.50 3568.75 1250.0 7487.50 1250.0 3562.50 7481.25 Mean Difference = Calibration Error% of Span= CEM # ---'-1---=----------- Span: 0-10000 PPM Zero/Low 1 Mid 1 High 1 5.17 42.50 9.58 3 3 3 .05 .42 .10 Test Times (start and finish): 13:54 --14:04 Cylinder ID No. CC266951 Certification Date : 4 Oct 2022 Expiration Date: 4 Oct 2030 Cylinder ID No. CC213578 Certification Date: 1 Sep 2023 Expiration Date: 1 Sep 2031 Cylinder ID No. CC330683 Certification Date : 1 Sep 2023 Expiration Date: 1 Sep 2031 Comments: 1 Absolute Difference= I Calibration value - Monitor Response I 2 Mean Difference = Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Calibration Error Test Field Data Sheet Date of Test: 9 May 2024 Quarter: Monitor Type: High CO Manufacturer: Servomex ------------Serial Number: 653305 ------------ Run Number Calibration mA Value Value 1 -zero 1251. 6.08 2-mid 3520. 9.70 3 -hi h 7495 . 15.99 4-mid 1251. 9.70 5 -zero 3520. 6.ld' 6-hi h 7495. 15.97 7 -zero 1251. 6.09 8-mid 3520. 9.69 9-hi h 7495. 15.98 Test Technicians: Nathan Haney _____ __,__ _____ _ Monitor Response 1300.0 3562.50 7493.75 3562.50 13'12.50 7481.25 1306.25 3556.25 7487 .50 Mean Difference = Calibration Error% of Span= CEM # _2~---------- Span: 0 -10000 PPM Zero/Low 1 Mid 1 High 1 55.25 40.42 7.50 3 3 3 .55 .40 .08 Test Times (start and finish): 15:08 --15:17 Cylinder ID No. CC266951 Certification Date: 4 Oct 2022 Expiration Date: 4 Oct 2030 Cylinder ID No . CC213578 Certification Date: 1 Sep 2023 Expiration Date: 1 Sep 2031 Cylinder ID No. CC330683 Certification Date: 1 Sep 202 3 Expiration Date: 1 Sep 2031 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference= Sum of Absolute Differences/ 3 3 Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fa ii </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: 16 May 2024 Quarter: J., l, Monitor Type: 02 Manufacturer: Servomex Serial Number: 4021 Run Number Calibration Value I -zero .996 2-mid 9.033 3 -hi h 15.04 4-mid .996 5 -zero 9.033 6-hi h 15.04 7 -zero .996 8-mid 9.033 15.04 '-.,UllllllUVU::> .i..:,1111:,::,1Ull lV.lVlllLVl Calibration Error Test Field Data Sheet Test Technicians : Nathan Haney ------''------- mA Value 4.63 9.78 13.65 9.80 4.64 13.64 4.64 9.78 13.64 Monitor Response .98 9.03 15.08 9.06 1.0 15.06 1.0 9.03 15.06 Mean Difference = Calibration Error% of Span= CEM # --=-1 __________ _ Span: 0-25 % Zero/Low 1 Mid 1 High 1 .01 .01 .03 3 3 3 .03 .04 .11 Test Times (start and finish): 13:37 --13:48 Cylinder ID No. EB0122991 Certification Date: 1 Nov 2021 Expiration Date: 1 Nov 2029 Cylinder ID No. CC47205 Certification Date: 14Jul 2020 Expiration Date: 14Jul2028 Cylinder ID No. CC151392 Certification Date: 31 Aug 2020 Expiration Date: 31 Aug 2028 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference = Sum of Absolute Differences/ 3 3 Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% ~Vl1Lll1UUU;:) .L,11J.1;:);:)1Vl1 H.1Vl1HV1 Calibration Error Test Field Data Sheet Date of Test: 16 May 2024 Test Technicians: Nathan Haney _____ ...,__ _____ _ Quarter: Monitor Type: 02 Manufacturer: Servomex ------------Serial Number: 652525 ------------ Run Number Calibration mA Value Value 1 -zero .996 4.59 2-mid 9.033 9.75 3 -hi h 15.04 13.62 4-mid .996 9.73 5 -zero 9.033 '4.60 6-hi h 15.04 13.63 7 -zero .996 4.61 8-mid 9.033 9.77 9-hi h 15.04 13.64 Monitor Response .92 8.98 15.03 8.95 .94 15.05 .95 9.02 15.06 Mean Difference = Calibration Error% of Span= CEM# -=-2 __________ _ Span: 0-25 % Zero/Low 1 Mid 1 High 1 2 2 2 .06 .OS .01 3 3 3 .23 .19 .OS Test Times (start and finish): 13:20 -13:35 Cylinder ID No. EB0122991 Certification Date: 1 Nov 2021 Expiration Date: 1 Nov 2029 Cylinder ID No. CC47205 Certification Date: 14Jul 2020 Expiration Date: 14Jul2028 Cylinder ID No. CC151392 Certification Date : 31 Aug 2020 Expiration Date: 31 Aug 2028 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference = Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: Quarter: Monitor Type: Manufacturer: Serial Number: 10 May 2024 THC Thermo Scientific 1203299855 Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney _____ _,__ _____ _ CEM# -----=-1 __________ _ ------------Span: 0 -100 PPM Run Number Calibration mA Value Value 1 -zero .0 4.0 2-mid 35.04 9.52 3 -hi h 75.38 15.87 4-mid .0 9.59 5 -zero 35.04 4.02 6-hi h 75.38 15.92 7 -zero .0 4.03 8-mid 35.04 9.55 9-hi h 75.38 15.98 Monitor Response .00 34.50 74.19 34.94 .12 74.50 .19 34.69 74.88 Mean Difference = Calibration Error% of Span= Zero/Low1 Mid 1 High 1 .10 .33 .86 3 3 .10 .33 .86 Test Times (start and finish): 14:48 --15:07 Cylinder ID No. ALM024524 Certification Date: 24 Feb 202 4 Expiration Date: 24 Feb 2032 Cylinder ID No. EB0122971 Certification Date: 3 Oct 2022 Expiration Date: 3 Oct 2030 Cylinder ID No. CC189252 Certification Date : 6 Sep 2023 Expiration Date: 6 Sep 2031 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference= Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% 3 Date of Test: 15 May 2024 Quarter: ~ l- Monitor Type: THC Manufacturer: Thermo Scientific Serial Number: 1203299856 Run Number Calibration Value 1 -zero .0 2-mid 35.04 3 -hi h 75.38 4-mid .0 S -zero 35.04 6-hi h 75.38 7 -zero .0 8-mid 35.04 9-hi h 75.38 Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney _____ _,__ _____ _ mA Value 3.99 9.60 16.13 9.70 3.99 16.18 3.98 9.66 16.22 Monitor Response -.06 35.0 75.81 35.63 -.06 76.13 -.13 35.38 76.38 Mean Difference= Calibration Error% of Span= CEM # -'2=------------- Span: 0 -100 PPM Zero/Low 1 Mid 1 High 1 .08 .32 .72 3 3 3 .08 .32 .72 Test Times (start and finish): 14:14 --14:30 Cylinder ID No. ALM024524 Certification Date: 24 Feb 2024 Expiration Date: 24 Feb 2032 Cylinder ID No. EB0122971 Certification Date: 3 Oct 2022 Expiration Date: 3 Oct 2030 Cylinder ID No. CC189252 Certification Date: 6 Sep 2023 Expiration Date: 6 Sep 2031 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference = Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: 16 May 2024 Quarter: ~~ Monitor Type: Co2 Manufacturer: Servomex Serial Number: 653204 Run Number Calibration Value 1 -zero 2.007 2 -mid 7.006 3 -hi 15.14 4-mid 2.007 5 -zero 7.006 6-hi h 15.14 7 -zero 2.007 8-mid 7.006 _J -high 15.14 \..,UilllilUUUS .CIIU:SSlUil 1v1unuur Calibration Error Test Field Data Sheet Test Technicians : Nathan Haney _____ .,__ _____ _ mA Value 5.76 9.68 16.12 9.76 5.76 16.08 5.77 9.72 16.10 Monitor Response 2.20 7.10 15.15 7.20 2.20 15.10 2.21 7.15 15.13 Mean Difference = Calibration Error% of Span= CEM # --=--1 __________ _ Span: 0-20 % Zero/Low 1 Mid 1 High 1 .20 .14 .02 3 3 3 .99 .72 .11 Test Times (start and finish): 10:33 --10:43 Cylinder ID No. XC0313108 Certification Date: 30 Aug 2016 Expiration Date: 30Aug 2024 Cylinder ID No. EB0123063 Certification Date: 23 Dec 2019 Expiration Date: 23 Dec 2027 Cylinder ID No. EB0106597 Certification Date: 5 Mar 2019 Expiration Date: 5 Mar 2027 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference = Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5% Date of Test: 16 May 2024 Quarter: ¥ ;x. Monitor Type: Co2 Manufacturer: Servomex Serial Number: 653305 Run Number Calibration Value 1 -zero 2.007 2-mid 7.006 3 -hi h 15.14 4-mid 2.007 5 -zero 7.006 6-hi h 15.14 7 -zero 2.007 8-mid 7.006 9 -pigh 15.14 ------· vUULlllUUU~ Dllll:):)1U11 J.V1U11UU1 Calibration Error Test Field Data Sheet Test Technicians: Nathan Haney ------'-------- mA Value 5.77 9.71 16.13 9.73 5.76 16.10 5.77 9.70 Monitor Response 2.21 7.14 15.16 7.16 2.20 15.13 2.21 7.13 Mean Difference = Calibration Error% of Span= CEM # _2=------------ Span: 0-20 % Zero/Low 1 Mid 1 High 1 2 2 2 .20 .14 .01 3 3 3 1.01 .68 .07 Test Times (start and finish): 12:40 -12:50 Cylinder ID No. XC031310B Certification Date: 30 Aug 2016 Expiration Date: 30 Aug 2024 Cylinder ID No. EB0123063 Certification Date: 23 Dec 2019 Expiration Date: 23 Dec 2027 Cylinder ID No. EB0106597 Certification Date: 5 Mar 2019 Expiration Date: 5 Mar 2027 Comments: 1 Absolute Difference = I Calibration value - Monitor Response I 2 Mean Difference= Sum of Absolute Differences/ 3 3Calibration Error= (Mean Difference/ Span) x 100% 3 CO pass/fail </= 5% of span </=10 ppm (lo range) </=100 ppm(hi range) 2 02 and Co2 pass/fail Criteria Co2 and 02 </= 0.5%