Loading...
HomeMy WebLinkAboutDAQ-2024-0052281 DAQC-222-24 CAERS ID 10435 (B4) MEMORANDUM TO: STACK TEST FILE – STERIGENICS US, LLC - Ethylene Oxide Commercial Sterilization Plant – Salt Lake County THROUGH: Rik Ombach, Minor Source Oil and Gas Compliance Section Manager FROM: Kyle Greenberg, Environmental Scientist DATE: March 6, 2024 SUBJECT: Sources: Ethylene oxide emission-control system - Scrubber Contact: Daniel May: 262-930-5372 Location: 4201 West 700 South, Salt Lake City, UT 84104 Test Contractor: ECSI, Inc. Permit/AO#: DAQE-AN104350030-21 dated March 25, 2021 Action Code: TR Subject: Review of Stack Test Report dated November 3, 2023 On November 3, 2023, DAQ received a test report for the above listed unit. Testing was performed October 5, 2023, to demonstrate compliance with the emission limits found in condition II.B.3.a of Approval Order DAQE-AN104350030-21. The calculated test results are: Source Test Date Test Method Pollutant Tester Results DAQ Results Limits Ethylene oxide emission- control system – Scrubber Backvents Oct. 5, 2023 CARB Method 431 ethylene oxide 99.953% Reduction Efficiency 99.972% Reduction Efficiency 99% or Greater Reduction Efficiency Ethylene oxide emission- control system – Scrubber Aeration Cells CARB Method 431 ethylene oxide 99.946% Reduction Efficiency 99.946% Reduction Efficiency 99% or Greater Reduction Efficiency DEVIATIONS: None. CONCLUSION: The stack test report appears to be acceptable. RECOMMENDATION: It’s recommended the emissions from the Ethylene oxide emission- control system be considered to have been in compliance with the emission limits of the Approval Order, during the time of testing. ATTACHMENTS: DAQ stack test review Excel Spreadsheets; Sterigenics US’ Test Report. Sterigenics US, LLC Sterigenics US, LLC Ethylene Oxide Commercial Sterilization Plant Ethylene Oxide Commercial Sterilization Plant Salt Lake County, Utah Salt Lake County, Utah Scrubber - Backvents Scrubber - Backvents Test Date: 10/5/2023 Test Date: 10/5/2023 Run #1 2 3 Average Run #1 2 3 Average Start Time 9:13 10:40 13:05 Start Time 9:13 10:40 13:05 Stop Time 9:28 10:55 13:20 Stop Time 9:28 10:55 13:20 Sample Duration (minutes)15 15 15 Sample Duration (minutes)15 15 15 wet C2H4O (ppmvw)0.0062 0.0062 0.0062 0.0062 wet C2H4O (ppmvw)14.350 23.9 27.5 21.9 Run #1 2 3 Average Run #1 2 3 Average Bws Moisture Content (%/100)0.000 0.000 0.000 0.000 Bws Moisture Content (%/100)0.000 0.000 0.000 0.000 dry C2H4O (ppmvd)0.0 0.0 0.0 0.0 dry C2H4O (ppmvd)14.4 23.9 27.5 21.9 1 2 3 Average Permit Limits dry Outlet C2H4O (ppmvd)0.0062 0.0062 0.0062 0.0062 dry Inlet C2H4O (ppmvd)14.3500 23.8750 27.5008 21.9086 DRE % Destruction Efficancy C2H4O (ppmvd)99.957%99.974%99.977%99.972%99% Reference Method Calculations Field Reference Method Data (INLET) DRE Calculations Field Reference Method Data (OUTLET) Reference Method Calculations Sterigenics US, LLC Sterigenics US, LLC Ethylene Oxide Commercial Sterilization Plant Ethylene Oxide Commercial Sterilization Plant Salt Lake County, Utah Salt Lake County, Utah Scrubber - Aeration Cells Scrubber - Aeration Cells Test Date: 10/5/2023 Test Date: 10/5/2023 Run #1 2 3 Average Run #1 2 3 Average Start Time 9:28 10:58 11:58 Start Time 9:28 10:58 11:58 Stop Time 10:28 11:58 12:58 Stop Time 10:28 11:58 12:58 Sample Duration (minutes)60 60 60 Sample Duration (minutes)60 60 60 wet C2H4O (ppmvw)0.0062 0.0062 0.0062 0.0062 wet C2H4O (ppmvw)10.089 11.925 12.733 11.6 Run #1 2 3 Average Run #1 2 3 Average Bws Moisture Content (%/100)0.000 0.000 0.000 0.000 Bws Moisture Content (%/100)0.000 0.000 0.000 0.000 dry C2H4O (ppmvd)0.0 0.0 0.0 0.0 dry C2H4O (ppmvd)10.1 11.9 12.7 11.6 1 2 3 Average Permit Limits dry Outlet C2H4O (ppmvd)0.0062 0.0062 0.0062 0.0062 dry Inlet C2H4O (ppmvd)10.0892 11.9250 12.7333 11.5825 DRE % Destruction Efficancy C2H4O (ppmvd)99.939%99.948%99.951%99.946%99% Reference Method Calculations Field Reference Method Data (INLET) DRE Calculations Field Reference Method Data (OUTLET) Reference Method Calculations ECSi REPORT OF AIR POLLUTION SOURCE TESTING OF AN ETHYLENE OXIDE EMISSION-CONTROL SYSTEM OPERATED BY STERIGENICS U.S., LLC. IN SALT LAKE CITY, UTAH ON OCTOBER 5, 2023 Submitted to: UTAH DEPARTMENT OF ENVIRONMENTAL QUALITY Division of Air Quality 150 North 1950 West Salt Lake City, Utah 84114-4820 Submitted by: STERIGENICS, LLC. 5725 West Harold Gatty Drive Salt Lake City, Utah 84116 Prepared by: ECSI, INC. PO Box 1498 San Clemente, California 92674-1498 November 3, 2023 ECSi i CONTACT SUMMARY CLIENT FACILITY Mr. Daniel May Mr. Joseph Jeppson Director, EH&S - EO Americas General Manager STERIGENICS US, LLC. STERIGENICS US, LLC. 2015 Spring Road, Suite 650 5725 West Harold Gatty Drive Oak Brook, Illinois 60523 Salt Lake City, Utah 84116 Phone: (262)930-5372 Phone: (801)328-9901 Email: dmay2@sterigenics.com Email: jjeppson@sterigenics.com TEST DATE Thursday, October 5, 2023 REGULATORY AGENCY Mr. Chad Gilgen Environmental Scientist, Minor Source Compliance UTAH DEPARTMENT OF ENVIRONMENTAL QUALITY (UDEQ) Department of Air Quality 150 North 1950 West Salt Lake City, Utah 84114-4820 Phone: (801)536-4237 Email: cgilgen@utah.gov TESTING CONTRACTOR Daniel P. Kremer President ECSi, Inc. PO Box 1498 San Clemente, California 92674-1498 Phone: (949)400-9145 email: dankremer@ecsi1.com website: www.ecsi1.com I hereby verify that ECSi is an independent testing laboratory that meets all of the criteria outlined in SCAQMD Rule 304(k), and that I personally review and certify all test results reported. I have performed over 3,000 CARB Method 431 ethylene oxide source tests since 1992. Daniel P. Kremer ECSi ii TABLE OF CONTENTS PAGE NO. CONTACT SUMMARY i TABLE OF CONTENTS ii LIST OF TABLES iii LIST OF APPENDICES iv 1.0 INTRODUCTION 1 2.0 EQUIPMENT 2 3.0 TESTING 3 4.0 RULE/COMPLIANCE REQUIREMENTS 4 5.0 TEST METHOD REFERENCE 5 5.1 Summary/Introduction 5 5.2 EtO Control Efficiency Measurement 5 5.3 Sample Transport 6 5.4 GC Injection 6 5.5 GC Conditions 7 5.6 Calibration Standards 7 5.7 Sampling Duration 8 5.8 Control Efficiency Calculations 8 6.0 TEST SCENARIO 9 7.0 QA/QC 10 7.1 Field Testing Quality Assurance 10 7.2 Calibration Procedures 10 8.0 TEST RESULTS 11 TABLES 12 APPENDICES 15 ECSi iii LIST OF TABLES TABLE DESCRIPTION PAGE NO. 1 Ethylene Oxide Control Efficiency – Backvent 13 2 Ethylene Oxide Control Efficiency –Aeration 14 ECSi iv LIST OF APPENDICES APPENDIX DESCRIPTION PAGE NO. A Calibration Data A-1 B Run #1 Chromatograms - Backvent B-1 C Run #1 Chromatograms - Aeration C-1 D Run #2 Chromatograms - Backvent D-1 E Run #2 Chromatograms - Aeration E-1 F Run #3 Chromatograms - Aeration F-1 G Run #3 Chromatograms - Backvent G-1 H Calibration Gas Certificates H-1 ECSi 1 1.0 INTRODUCTION On Thursday, October 5, 2023, ECSi performed air pollution source testing of an ethylene oxide (EtO) emission-control system operated by Sterigenics U.S., LLC. in Salt Lake City, Utah. The control device tested was a two-stage Advanced Air Technologies Safe Cell emission-control system, which is currently used to control emissions from fourteen EtO aeration cells. The purpose of the testing program was to demonstrate continued compliance with the conditions established in the Air Quality Permit granted to Sterigenics by the Utah Department of Environmental Quality (UDEQ). ECSi 2 2.0 EQUIPMENT The gas-sterilization system is comprised of ten commercial sterilizers, which are discharged through liquid- ring vacuum pumps to a Ceilcote packed tower scrubber emission-control system, ten sterilizer exhaust vents (backvents), which were discharged to atmosphere at the time of the test, and fourteen aeration cells, which are discharged to an existing two-stage Advanced Air Technologies (AAT) Safe Cell emission-control system. As an alternative emission-control scenario, the facility also has the capability to discharge the outlet of the Ceilcote scrubber to the inlet of the AAT Safe Cell system, referred to as a “double scrub” configuration, for maximum emissions reduction. The gas-sterilization and emission-control equipment consist of the following: • Six Vacudyne Gas Sterilizers, all Model 810, each comprised of a steam-heated 795 cubic foot interior volume sterilization chamber, a recirculating vacuum pump chamber evacuation system, a backdraft valve, and a fugitive emissions exhaust hood; • One Vacudyne Gas Sterilizer comprised of a steam-heated 3600 cubic foot interior volume sterilization chamber, a recirculating vacuum pump chamber evacuation system, a backdraft valve, and a fugitive emissions exhaust hood; • One American Sterilizer Company Gas Sterilizer, Model 1200, comprised of a steam-heated 1133 cubic foot interior volume sterilization chamber, a recirculating vacuum pump chamber evacuation system, a backdraft valve, and a fugitive emissions exhaust hood; • One Environmental Tectonics Corporation Gas Sterilizer, Model 1035, comprised of a steam-heated 283 cubic foot interior volume sterilization chamber, a recirculating vacuum pump chamber evacuation system, a backdraft valve, and a fugitive emissions exhaust hood; • One National Sterilizer Company Gas Sterilizer, comprised of a steam-heated 35 cubic foot interior volume sterilization chamber, a recirculating vacuum pump chamber evacuation system, a backdraft valve, and a fugitive emissions exhaust hood • Fourteen Aeration Chambers, each comprised of a heated aeration chamber and a chamber exhaust system. ECSi 3 Sterilizer vacuum pump emissions are be controlled by: • One Ceilcote packed tower chemical scrubber, equipped with: a reaction/interface column, 29’ 4” high, 48” in diameter, with a 20’ bed of #1 Tellerette packing; a 150 GPM scrubber fluid recirculation system; and two 17,000-gallon reaction/storage tanks. Aeration emissions are controlled by: One two-stage Advanced Air Technologies Safe Cell emission-control system, comprised of a packed-tower chemical scrubber (SC1), equipped with a packed reaction/interface column, a scrubber fluid recirculation system, and a scrubber fluid reaction/storage tank, and a dry bed reactor/scrubber (SC2), comprised of a bank of solid-bed reaction vessels, connected in parallel, installed downstream of SC1 and upstream of a dedicated blower system. ECSi 4 3.0 TESTING EtO source testing was conducted in accordance with the procedures outlined in CARB Method 431, the USEPA approved method specified in 40 CFR, Part 63.365, subpart O. EtO concentration measurement for each test run was conducted simultaneously at the inlet and outlet of the AAT Safe Cell System during chamber backvent, and during a one-hour interval of the 24-hour aeration process. A total of three chamber backvent test runs, and three one-hour aeration test runs, were performed. During backvent and aeration testing, EtO concentration at the inlet and the outlet of the AAT Safe Cell System was determined using direct source sample injection into the gas chromatograph (GC). All backvent and aeration testing was performed using freshly sterilized product. The testing program was conducted in accordance with the procedures outlined in the following sections. ECSi 5 4.0 RULE/COMPLIANCE REQUIREMENTS The EtO gas-sterilization system at Sterigenics U.S., LLC. was tested to evaluate compliance with the requirements specified in the TCEQ Permit. The current testing was performed to demonstrate continued compliance with the following requirement: • Aeration and backvent emissions must be discharged to control equipment which achieves an EtO emission-reduction efficiency of at least 99.0%. Testing is required to demonstrate compliance with these requirements. Source testing of the AAT Safe Cell System is required once every 5 years. ECSi 6 5.0 TEST METHOD REFERENCE 5.1 INTRODUCTION EtO source testing was conducted in accordance with the procedures outlined in CARB Method 431, the USEPA approved method specified in 40 CFR, Part 63.365, subpart O. EtO concentration measurement for each test run was conducted simultaneously at the inlet and outlet of the AAT Safe Cell System during chamber backvent, and during a one-hour interval of the 24-hour aeration process. A total of three chamber backvent test runs, and three one-hour aeration test runs, were performed. During backvent and aeration testing, EtO concentration at the inlet and the outlet of the AAT Safe Cell System were determined using direct source sample injection into the gas chromatograph (GC). All backvent and aeration testing was performed using freshly sterilized product. Operation and documentation of process conditions was performed by personnel from Sterigenics U.S., LLC. using existing monitoring instruments installed by the manufacturer of the equipment to be tested. In accordance with TCEQ requirements, and the procedures established in USEPA 40 CFR, Part 63.365, Subpart O, the following parameter was recorded: scrubber liquor level. 5.2 CONTROL EFFICIENCY MEASUREMENT During backvent and aeration testing, EtO concentration at the inlet and outlet of the AAT Safe Cell System was determined using direct source sample injection into the GC. Since the source gas flow is identical at the inlet and outlet of the AAT Safe Cell System control-efficiency of EtO during aeration and backvent was calculated by comparing the concentration of EtO vented to the system inlet to the concentration of EtO vented from the system outlet. CARB Method 431, Appendix A, specifies that AAT Safe Cell System emission-control devices may be tested, and control efficiency determined, without volumetric flow measurement as long as the following criteria are met: 1) There is no dilution between the inlet and outlet sampling locations 2) There is identical flow at the inlet and outlet sampling locations, and 3) There is constant flow throughout the duration of the compliance test. ECSi 7 These conditions were all met during the testing performed at Sterigenics. There is no dilution, and no introduction of any outside source of airflow into the AAT system between the inlet and outlet sampling ports. Further, prior years’ testing has exhaustively proven that the inlet and outlet volumetric flows are identical. During backvent and aeration, vented gas was analyzed by an SRI, Model 8610, portable gas chromatograph (GC), equipped with the following: dual, heated sample loops and injectors; dual columns; and dual detectors. A flame ionization detector (FID) was used to quantify inlet EtO concentration, and a photoionization detector (PID) was used to quantify low-level EtO concentration at the emission-control device outlet. 5.3 SAMPLE TRANSPORT Source gas was pumped to the GC at approximately 2000 cubic centimeters per minute (cc/min) from the sampling ports through two lengths of Teflon® sample line, each with a nominal volume of approximately 75 cubic centimeters (cc) and an outer diameter of 0.25 inch. At the inlet, the sampling port was located in the common backvent/aeration discharge duct, upstream of the oxidizer. At the outlet of the AAT Safe Cell System, sampling ports were located in the exhaust stack downstream of the catalyst bed. 5.4 GC INJECTION Source-gas samples were then injected into the GC which was equipped with two heated sampling loops, each containing a volume of approximately 2cc and maintained at 100 degrees Celsius (C). Injections occurred at approximately five-minute intervals during aeration testing, and at approximately one-minute intervals during the backvent testing. Helium was the carrier gas for both the FID and PID. 5.5 GC CONDITIONS The packed columns for the GC were both operated at 90 degrees C. The columns were stainless steel, 6 feet long, 0.125-inch outer diameter, packed with 1 percent SP-1000 on 60/80 mesh Carbopack B. During the analysis, the FID was operated at 250 degrees C. The support gases for the FID were helium (99.999% pure), hydrogen (99.995% pure) and air (99.9999% pure). Any unused sample gas was vented from the GC system back to the inlet of the control device being tested. ECSi 8 5.6 CALIBRATION STANDARDS The FID was calibrated for mid-range part-per-million-by-volume (ppmv) level analysis using gas proportions similar to the following: 1) 100 ppmv EtO, balance nitrogen 2) 50 ppmv EtO, balance nitrogen (audit gas) 3) 10 ppmv EtO, balance nitrogen 4) 1 ppmv EtO, balance nitrogen The PID was calibrated for low-range ppmv level analyses using gas proportions similar to the following: 1) 100 ppmv EtO, balance nitrogen 2) 50 ppmv EtO, balance nitrogen (audit gas) 3) 10 ppmv EtO, balance nitrogen 4) 1 ppmv EtO, balance nitrogen Each of these calibration standards was in a separate, certified manufacturer's cylinder. Copies of the calibration gas laboratory certificates are attached as Appendix H. 5.7 SAMPLING DURATION Backvent testing was performed in conjunction with normal production operations, during the chamber exhaust venting which is conducted for each sterilization chamber upon conclusion of the sterilization cycle, immediately prior to and during chamber unloading. Backvent sampling duration was 15 minutes for each of the three test runs. Since aeration is a 24-hour process at this facility, with constant discharge flow from the aeration chambers to the emission-control system, aeration testing consisted of three 1-hour test runs. Each test run was performed with freshly sterilized product in the aeration chambers. ECSi 9 5.8 CONTROL-EFFICIENCY CALCULATIONS Control efficiency of EtO was calculated for aeration and backvent, using the following CARB-approved equation: Efficiency = (Ci - Co / Ci)(100) Which is a mathematical simplification of the following equation from CARB Method 431, with the identical inlet/outlet flow value removed: Efficiency = (Wi - Wo / Wi)(100) Where: Wi = Mass flow rate to the control device inlet, pounds, calculated as (Ci)(Fi) Where: Ci = EtO concentration at the control device inlet Fi = Flow rate at the control device inlet Wo = Mass flow rate from the control device outlet, pounds, calculated as (Co)(Fo) Where: Co = EtO concentration at the control device outlet Fo = Flow rate at the control device outlet Results of the control-efficiency testing are presented in Section 8.0, and in Tables 1 and 2. ECSi 10 6.0 TEST SCENARIO The backvent and aeration testing was performed during normal process load conditions. Three backvent and three aeration test runs were conducted in series to verify the performance of the emission-control device. The testing schedule was as follows: 1) Testing equipment was set up and calibrated. 2) Backvent Test Run #1 was conducted with one freshly sterilized production load. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 3) Aeration Test Run #1 was conducted with freshly sterilized product in aeration. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 4) Backvent Test Run #2 was conducted with freshly sterilized product in aeration. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 5) Aeration Test Run #2 was conducted with one freshly sterilized production load. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 6) Aeration Test Run #3 was conducted with freshly sterilized product in aeration. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 7) Backvent Test Run #3 was conducted with one freshly sterilized production load. Sampling was performed at the inlet and the outlet of the AAT Safe Cell System. 8) Post calibration check was performed, testing equipment was packed. ECSi 11 7.0 QA/QC 7.1 FIELD TESTING QUALITY ASSURANCE At the beginning of the test, the sampling system was leak checked at a vacuum of 15 inches of mercury. The sampling system was considered leak free when the flow indicated by the rotameters fell to zero. At the beginning of the test, a system blank was analyzed to ensure that the sampling system was free of EtO. Ambient air was introduced at the end of the sample line and drawn through the sampling system line to the GC for analysis. The resulting chromatogram also provided a background level for non-EtO components (i.e. ambient air, carbon dioxide, water vapor) which are present in the source gas stream due to the ambient dilution air which is drawn into the emission-control device, and due to the destruction of EtO by the emission-control device which produces carbon dioxide and water vapor. This chromatogram, designated AMB, is included with the calibration data in Appendix A. 7.2 CALIBRATION PROCEDURES The GC system was calibrated prior to testing. The pre-test calibration procedure included triplicate injections of each concentration of calibration gas, for each detector. The lowest concentration of calibration gas was injected 7-10 times for the detector used at the outlet sampling point, as part of the method detection limit (MDL) determination for the test. The MDL calculations were performed using a spreadsheet provided to ECSi by Ned Shappley of the USEPA Measurement Technology Group, in accordance with USEPA Method 301. The resulting calibration data was entered into the Peaksimple II analytical software, and a calibration curve for the test was established for each detector. A gas cylinder of similar composition as the calibration gases, but certified by a separate supplier, was used to verify calibration gas composition and GC performance (audit gas). This gas was used as a calibration check at the test’s midpoint, and at the test’s conclusion, to verify that the Peaksimple calibration curve for each detector was still accurate within 10% of the mean values established in the multipoint calibration. All calibration gases and support gases used were of the highest purity and quality available. A copy of the laboratory certification for each calibration gas is attached as Appendix H. ECSi 12 8.0 TEST RESULTS The AAT Safe Cell System was found to have an average EtO control efficiency of 99.953 percent for backvent, and an average EtO control efficiency of 99.946 percent for aeration. In accordance with state and federal requirements, backvent and aeration discharge streams must be vented to control equipment with an EtO emission-reduction efficiency of at least 99 percent. The AAT Safe Cell System met this requirement. The test results are summarized in Tables 1 and 2. These tables include results for EtO control efficiency of the emission-control device. Chromatograms and chromatographic supporting data are attached as Appendices A through G. ECSi 13 TABLES TABLE 1 ETHYLENE OXIDE CONTROL EFFICIENCY - BACKVENT OF AN AAT SAFE CELL ETHYLENE OXIDE EMISSION CONTROL SYSTEM OPERATED BY STERIGENICS, INC. IN SALT LAKE CITY, UTAH ON OCTOBER 5, 2023 RUN INJECTION INLET ETO OUTLET ETO ETO CONTROL NUMBER TIME CONC. (PPM)(1)CONC. (PPM)(2)EFFICIENCY 1(3) 914 80.2 0.0062 99.992 1 915 9.67 0.0062 99.936 1 916 9.34 0.0062 99.934 1 918 8.90 0.0062 99.930 1 919 8.83 0.0062 99.930 1 920 8.87 0.0062 99.930 1 921 8.70 0.0062 99.929 1 922 8.78 0.0062 99.929 1 924 8.65 0.0062 99.928 1 925 8.70 0.0062 99.929 1 926 8.60 0.0062 99.928 1 927 8.79 0.0062 99.929 1 928 8.52 0.0062 99.927 2(4) 1041 134 0.0062 99.995 2 1043 23.5 0.0062 99.974 2 1044 14.1 0.0062 99.956 2 1045 13.2 0.0062 99.953 2 1046 13.4 0.0062 99.954 2 1047 13.0 0.0062 99.952 2 1048 13.1 0.0062 99.953 2 1049 12.8 0.0062 99.952 2 1050 12.7 0.0062 99.951 2 1052 11.8 0.0062 99.947 2 1053 12.5 0.0062 99.950 2 1054 12.4 0.0062 99.950 3(5) 1306 9.41 0.0062 99.934 3 1307 124 0.0062 99.995 3 1309 25.8 0.0062 99.976 3 1310 22.3 0.0062 99.972 3 1311 21.3 0.0062 99.971 3 1312 19.8 0.0062 99.969 3 1313 19.1 0.0062 99.968 3 1314 18.1 0.0062 99.966 3 1315 17.3 0.0062 99.964 3 1317 18.2 0.0062 99.966 3 1318 17.1 0.0062 99.964 3 1319 17.6 0.0062 99.965 TIME-WEIGHTED AVERAGE: 21.70 0.0062 99.953 UDEQ REQUIRED CONTROL EFFICIENCY: 99% Notes: (1) - PPM = parts per million by volume (2) - 0.0062 ppm was the method detection limit on the day of the test, results with this value are reported as <0.0062 ppm. (3) - Backvent Phase Test Run #1 started at 09:13, ended at 09:28. (4) - Backvent Phase Test Run #2 started at 10:40, ended at 10:55. (5) - Backvent Phase Test Run #3 started at 13:05, ended at 13:20. ECSi TABLE 2 ETHYLENE OXIDE CONTROL EFFICIENCY - AERATION OF AN AAT SAFE CELL ETHYLENE OXIDE EMISSION CONTROL SYSTEM OPERATED BY STERIGENICS, INC. IN SALT LAKE CITY, UTAH ON OCTOBER 5, 2023 RUN INJECTION INLET ETO OUTLET ETO ETO CONTROL NUMBER TIME CONC. (PPM)(1)CONC. (PPM)(2)EFFICIENCY 1(3) 930 8.98 0.0062 99.931 1 935 9.40 0.0062 99.934 1 940 9.97 0.0062 99.938 1 945 9.93 0.0062 99.938 1 950 9.89 0.0062 99.937 1 955 10.1 0.0062 99.939 1 1000 10.0 0.0062 99.938 1 1005 10.0 0.0062 99.938 1 1010 10.2 0.0062 99.939 1 1015 11.0 0.0062 99.944 1 1020 10.8 0.0062 99.943 1 1025 10.8 0.0062 99.943 2(4) 1100 10.9 0.0062 99.943 2 1105 10.9 0.0062 99.943 2 1110 10.0 0.0062 99.938 2 1115 11.4 0.0062 99.946 2 1120 12.1 0.0062 99.949 2 1125 12.6 0.0062 99.951 2 1130 12.1 0.0062 99.949 2 1135 12.4 0.0062 99.950 2 1140 12.3 0.0062 99.950 2 1145 12.6 0.0062 99.951 2 1150 12.7 0.0062 99.951 2 1155 13.1 0.0062 99.953 3(5) 1200 13.4 0.0062 99.954 3 1205 13.5 0.0062 99.954 3 1210 13.1 0.0062 99.953 3 1215 13.3 0.0062 99.953 3 1220 13.3 0.0062 99.953 3 1225 13.2 0.0062 99.953 3 1230 12.2 0.0062 99.949 3 1235 12.4 0.0062 99.950 3 1240 12.2 0.0062 99.949 3 1245 12.0 0.0062 99.948 3 1250 12.2 0.0062 99.949 3 1255 12.0 0.0062 99.948 TIME-WEIGHTED AVERAGE: 11.58 0.0062 99.946 UDEQ REQUIRED CONTROL EFFICIENCY: 99% Notes: (1) - PPM = parts per million by volume (2) - 0.0062 ppm was the method detection limit on the day of the test, results with this value are reported as <0.0062 ppm. (3) - Aeration Phase Test Run #1 started at 09:28, ended at 10:28. (4) - Aeration Phase Test Run #2 started at 10:58, ended at 11:58. (5) - Aeration Phase Test Run #3 started at 11:58, ended at 12:58.ECSi ECSi 14 APPENDICES ECSi A-1 APPENDIX A Calibration Data ECSi B-1 APPENDIX B Run #1 Chromatograms - Backvent ECSi C-1 APPENDIX C Run #1 Chromatograms - Aeration ECSi D-1 APPENDIX D Run #2 Chromatograms - Backvent ECSi E-1 APPENDIX E Run #2 Chromatograms - Aeration ECSi F-1 APPENDIX F Run #3 Chromatograms - Aeration ECSi G-1 APPENDIX G Run #3 Chromatograms - Backvent ECSi H-1 APPENDIX H Gas Certifications